Infinium
®
HD Assay Gemini, Manual Protocol (Post-AMP)
Experienced User Card
Catalog # WG-901-3002 Page 23 of 28
Part # 11311015 Rev. A
[ ] c. 250 μl ATM and incubate for 10 minutes.
[ ] d. 450 μl XC3 and incubate for 1 minute. Repeat once, and then
wait 5 minutes.
[ ] e. 250 μl STM and incubate for 10 minutes.
[ ] f. 450 μl XC3 and incubate for 1 minute. Repeat once, and then
wait 5 minutes.
[ ] g. 250 μl ATM and incubate for 10 minutes.
[ ] h. 450 μl XC3 and incubate for 1 minute. Repeat once, and then
wait 5 minutes.
[ ] i. 250 μl STM and incubate for 10 minutes.
[ ] j. 450 μl XC3 and incubate for 1 minute. Repeat once, and then
wait 5 minutes.
[ ] 3. Immediately remove the Flow-Through Chambers from the Chamber
Rack and place horizontally on a lab bench at room temperature.
Wash and Coat
[ ] 1. Dispense PB1 into a wash dish as follows, and then cover the dish:
For 8 Beadchips, dispense 310 ml
For 16 Beadchips, dispense 300 ml
For 24 BeadChips, dispense 285 ml
[ ] 2. Place the staining rack inside the wash dish. The locking arms and tab
should face
towards you.
[ ] 3. For each BeadChip:
Use the dismantling tool to remove the two metal clamps from the
Flow-Through Chamber.
Remove the glass back plate, the spacer, and then the BeadChip.
Immediately place each BeadChip into the staining rack that is in the
PB1 wash dish, with the barcode facing
away from you. Place half of
the BeadChips above the handle and half below. All chips should b e
completely submerged.
[ ] 4. Slowly move the staining rack up and down 10 times, breaking the
surface of the reagent.
[ ] 5. Soak for 5 minutes.
[ ] 6. Dispense XC4 into a wash dish as follows:
For 8 Beadchips, dispense 310 ml
For 16 Beadchips, dispense 300 ml
For 24 BeadChips, dispense 285 ml
CAUTION
Handle the BeadChips only by the edges or the barcode
end. Do not let the BeadChips dry out.
CAUTION
Do not leave the BeadChips in the PB1 for more than
30 minutes.
[ ]
[ ]
[ ]
[ ]