Infinium
®
HD Assay Gemini, Manual Protocol (Post-AMP)
Experienced User Card
Catalog # WG-901-3002 Page 18 of 28
Part # 11311015 Rev. A
[ ] 5. Remove the IntelliHyb seal from each BeadChip as follows:
[ ] a. Using powder-free gloved hands, hold the BeadChip in one
hand with your thumb and forefinger on the long edges of the
BeadChip. The BeadChip may also be held with the thumb and
forefinger on the short edges of the BeadChip. In either case
avoid contact with the sample inlets. The barcode should be
facing up and be closest to you, and the top side of the
BeadChip should be angled slightly away from you.
[ ] b. Remove the entire seal in a single, rapid motion by pulling it off
in a diagonal direction. Start with a corner on the barcode end
and pull with a continuous upward motion away from you and
towards the opposite corner on the top side of the BeadChip.
Do not stop and start the pulling action. Do not touch the
exposed active areas.
[ ] 6. Immediately and carefully slide each BeadChip into the wash rack one at
a time, making sure that the BeadChip is completely submerged in the
PB1.
[ ] 7. Repeat steps 5 and 6 until all BeadChips are transferred to the
submerged wash rack. The wash rack holds up to 8 BeadChips.
[ ] 8. Once all BeadChips are in the wash rack, move the wash rack up and
down for 1
minute, breaking the surface of the PB1 with gentle, slow
agitation.
[ ] 9. Move the wash rack to the other wash dish containing PB1. Make sure
the BeadChips are completely sub merged.
[ ] 10. Move the wash rack up and down for 1 minute, breaking the surface of
the PB1 with gentle, slow agitation.
[ ] 11. If you are processing more than 8 BeadChips:
[ ] a. Complete the steps in the next section, Assemble Flow-Through
Chambers, for the first eight BeadChips.
[ ] b. Place the assembled Flow-Through Chambers of the first eight
BeadChips on the lab bench in a horizontal position.
[ ] c. Repeat steps 3 through 11 from this section for any additional
BeadChips. Use new PB1 for each set of eight BeadChips.